Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Astro-Med ASC-902: CAL ADJ /5520,34401 DATE: 26-Aug-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 111 CONFIGURATION: HP 34401A CONFIGURATION: Fluke 5520A ============================================================================= # # Source: # Astro-Med Verification / Procedure for ASC-902 # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Procedure not formally tested with UUT. 1.001 DISP See procedure README file. 1.002 ASK- N P F V 1.003 ASK+ K X 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Turn unit ON. 1.005 DISP 1.005 DISP Warm-up time: 30min - REQUIRED 1.005 DISP Temperature: 23 +/-2 degC 1.005 DISP Humidity: 55 +/-10% 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Extension Card: Astro-Med part number 30776-001 1.008 HEAD SET-UP PROCEDURE 1.009 DISP Turn UUT OFF! 1.009 DISP 1.010 DISP Remove the right side cover of the ASC-902. 1.010 DISP 1.010 DISP Connect the extension card to the signal conditioner PCB 1.010 DISP and reinsert into the card cage. 1.011 DISP Set the ASC-902 switches to the following positions: 1.011 DISP 1.011 DISP Filter: OFF 1.011 DISP Zero Suppression: OFF 1.011 DISP Mode: P-P 1.011 DISP Coupling: DC 1.011 DISP Polarity: NON-INV 1.012 DISP Place the Negative meter lead to common input 1.012 DISP (black binding post on back of card cage). 1.012 DISP 1.012 DISP Turn UUT ON. 1.013 DISP Short the inputs together by placing a jumper between 1.013 DISP the positive, negative, and common (red, grey, black) 1.013 DISP binding posts on the back of the card cage. 1.014 HEAD INPUT NULL 1.015 DISP Connect the 34401A to the UUT as follows: 1.015 DISP [32] 34401A Input HI to UUT TP6 1.015 DISP [32] 34401A Input LO to signal ground 1.016 DISP Set the Attenuator/Range to 1mV/div. 1.017 MESS Adjust R14 for a 34401A reading of 0 +/-1mV. 1.018 34401 0mV S 2W 1.019 HEAD RANGE NULL 1.020 DISP Set the Attenutor/Range TO 1V/div. 1.021 MESS Adjust R46 for a 34401A reading of 0 +/-1mV. 1.022 34401 0mV S 2W 1.023 HEAD FINAL NULL 1.024 DISP Connect the 34401A to the UUT as follows: 1.024 DISP [32] 34401A Input HI to pin 2 of edge connector or the 1.024 DISP [32] BNC connector of the back card cage (signal out). 1.024 DISP [32] 34401A Input LO to signal ground 1.025 MESS Adjust R11 for a 34401A reading of 0 +/-5mV. 1.026 34401 0mV S 2W 1.027 HEAD RMS NULL 1.028 DISP Set the Attenuator/Range to OFF. 1.028 DISP Set the Mode to RMS. 1.029 MESS Adjust R26 for a 34401A reading of 0 +/-5mV. 1.030 34401 0mV S 2W 1.031 HEAD P-P GAIN 1.032 DISP Connect the 5520A to the UUT binding posts as follows: 1.032 DISP [32] 5520A NORMAL HI to UUT positive (red) 1.032 DISP [32] 5520A NORMAL LO to UUT negative (grey) 1.032 DISP 1.032 DISP Leave the 34401A connected to the UUT. 1.033 DISP Set Mode to P-P 1.033 DISP Set Attenuator/Range to 10mV/div. 1.034 5520 0.500V S 2W 1.035 MESS Adjust R9 for a 34401A reading of 0.500 +/-0.005V. 1.036 34401 500mV S 2W 1.037 5520 * S 1.038 HEAD RMS GAIN 1.039 DISP Set Mode to RMS 1.039 DISP Set Attenuator/Range to 10mV/div. 1.040 5520 1.000V S 2W 1.041 MESS Adjust R24 of a 34401A reading of 1.000 +/-0.005V. 1.042 34401 1.000V S 2W 1.043 END